Draping Steps
Start with down sheet, 1 cm below the 10x10.
Place the bottom blue towel 1 cm cranial to the down sheet.
Position the top blue towel.
Place the lateral blue towel on the opposite side first with 4 cm ioban, bringing the extra ioban down onto the down drape.
Place the other lateral blue towel with 4 cm ioban on the surgery side, using the extra ioban for the down drape.
Drape with a lap drape that has circumferential pockets for irrigation, going slightly lateral to the surgery side to allow for a transforaminal approach. If also performing an interlaminar approach, ensure the midline is still visible.
Cover the C-arm on the opposite side.
Throw off suctions towards the suction machine; there will be three with tips cut off and clamps.
Measure suctions by placing them into the most southern part of the drape so that X-ray does not see them; clamp to towel, ensuring suction holes are not covered.
Cut two small holes in the cranial part of the pocket drape to feed suctions on each side through; the C-armor suction does not go through a hole.
Place one suction in the C-armor and one in each side of the pocket drape (three total).
Create a pocket for the Triggerflex on the opposite side of the surgeon on the patient’s shoulder blade.
Secure suctions with the right white Velcro.
Secure the Triggerflex and/or drill with the left Velcro.
Measure the Triggerflex and drill by placing the ends into the southern white Velcro or on the Mayo stand.
Assemble the drill:
Lubricate the drill bit with mineral oil.
Put the handpiece in the open position.
Connect the handpiece to the motor.
Twist the handpiece until it locks into place.
Insert the drill bit.
Twist the handpiece to the closed position.
Twist the support tube to lock it down.
Press the foot pedal slowly to test the drill.
Attach the light cord.
Attach a stopcock to the southernmost outflow of the scope.
Attach irrigation to the stopcock.
Prepare the scope condom for transforaminal approach:
Cut a small hole in the top of the condom and place it on the base; it should extend about 1 inch past the connection for the black cord.
White balance the camera (hold 'W') and ensure the zoom is appropriate; check the image quality.
Ensure the laser is on the field with a towel over the cord.
Place the last special clamp on the Mayo stand.
Cover the C-arm with a shower cap.
If using a retractor:
Attach the retractor clamp onto the clamp on the bed.
Position the arm of the retractor with the flat face facing out; angle the second arm at 45 degrees.